R Series - Auto Manual Testing Excerpt - v14

2 min read 1 year ago
Published on Aug 05, 2024 This response is partially generated with the help of AI. It may contain inaccuracies.

Table of Contents

Introduction

This tutorial provides a step-by-step guide on performing manual testing for the ZOLL R Series monitor/defibrillator. Following these instructions ensures that the device is Code-Ready and compliant with current guidelines for resuscitation support.

Step 1: Verify Automatic Test Results

  • After the automatic self-test is completed, check the code readiness indicator window.
  • Look for:
    • A green checkmark: This indicates successful completion of the test.
    • A red X: This signifies that the R Series did not pass one of the automatic test requirements.
  • If you see a red X, take corrective action.

Step 2: Take Corrective Action

  • If a red X is present:
    • Replace expired electrodes.
    • Perform a manual test.
  • If the red X does not change to a green checkmark after these actions:
    • Contact clinical engineering for assistance in performing a manual test.

Step 3: Set Up for Manual Testing

  • Ensure the OneStep cable is connected to either:
    • OneStep electrodes
    • Paddles
    • Test port
  • Turn the mode selector to Defib.

Step 4: Configure Energy Settings

  • Set the energy to 30 joules.
  • Press Charge.
  • Wait for the charge ready tone to sound.

Step 5: Adjust Energy Level

  • Press the down energy select button until the energy level reaches 20 joules.
  • Note that the defibrillator will disarm itself at this point.

Step 6: Reset Energy Level

  • Press the energy select buttons to reset the energy back to 30 joules.
  • Press Charge again.

Step 7: Deliver Shock

  • Press Shock.
  • Check the display for the message Test OK.
  • If this message does not appear, notify clinical engineering for further assistance.

Step 8: Follow Institutional Policies

  • Review your institution's policy to confirm the frequency of manual testing required for the R Series monitor/defibrillator.

Conclusion

Regular testing and maintenance of the ZOLL R Series monitor/defibrillator are crucial for ensuring its reliability during emergencies. By following the steps outlined in this tutorial, you can effectively conduct manual tests and address any issues that arise. For ongoing compliance, always stay updated on your institution's policies regarding equipment testing frequency.